MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FT4 III; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FT4 III; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Model Number FT4 G3
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/26/2022
Event Type  malfunction  
Manufacturer Narrative
The investigation determined that the result differences generated between the different methods are consistent with methodological differences.This relates to the overall setups of the assays, the antibodies used, differences in reference materials/methods, and the standardized methodology used.Based on the available data, a general reagent issue could be excluded.The investigation did not identify a product problem.
 
Event Description
There was an allegation of questionable elecsys ft4 iii assay results for 1 patient sample on a cobas 8000 e 801 module compared to an abbott architect.The patient sample was sent to another lab for testing due to the normal tsh result accompanied by high ft3 and ft4 results.Refer to the attachment to the medwatch for all patient data.No questionable results were reported outside of the laboratory.The customer's analyzer serial number is (b)(4).The investigational analyzer serial number is (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15882296
MDR Text Key307799702
Report Number1823260-2022-03802
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Model NumberFT4 G3
Device Catalogue Number07976887190
Device Lot Number630888
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-