• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 SPIRIT® HYDROCOLLOID ADHESIVE SHEATH; MALE EXTERNAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 SPIRIT® HYDROCOLLOID ADHESIVE SHEATH; MALE EXTERNAL CATHETER Back to Search Results
Model Number 39305
Device Problem Nonstandard Device (1420)
Patient Problem Skin Inflammation/ Irritation (4545)
Event Date 11/14/2022
Event Type  malfunction  
Event Description
It was reported that the male external catheter was hard to remove due to all the adhesive that runs through the sheath.They stated that it sometimes takes the skin off the bottom of their penis.Per additional information received via liberator on 14-nov-2022, stated that these male external catheters pull at skin and made it soreness to the patient.No medical intervention was reported.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Manufacturer Narrative
The reported event was inconclusive as no sample was returned for evaluation.A potential root cause for this failure could be "operator error".It was unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "description/indication the self-adhering male external catheter is designed for the management of male urinary incontinence.Contraindication do not use on irritated or compromised skin.Precaution: do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply 1) wash penis with mild soap and warm water.Dry thoroughly.2) trim pubic hair if necessary.3) open package at perforation.4) to remove plastic insert, squeeze catheter at the top of the white cone and pull to release.5) unroll self-adhering catheter over penis.6) gently squeeze the catheter to properly seal adhesive to the skin.7) connect to collection bag.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event Description
It was reported that the male external catheter was hard to remove due to all the adhesive that runs through the sheath.They stated that it sometimes takes the skin off the bottom of their penis.Per additional information received via liberator on 14-nov-2022, stated that these male external catheters pull at skin and made it soreness to the patient.No medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPIRIT® HYDROCOLLOID ADHESIVE SHEATH
Type of Device
MALE EXTERNAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15887786
MDR Text Key307687428
Report Number1018233-2022-09123
Device Sequence Number1
Product Code NNX
UDI-Device Identifier00801741071362
UDI-Public(01)00801741071362
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number39305
Device Catalogue Number39305
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;