Model Number V60 VENT, CHINA OPT: CFLEX, AVAPS
Device Problem Volume Accuracy Problem (1675)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/03/2022
Event Type  malfunction  
Event Description
An abnormal tidal volume alarm is found during equipment self inspection.Investigation is ongoing.
 
Manufacturer Narrative
H10 it was confirmed that the system was not in clinical use; there was no reported harm to patient/user.Issue occurred during self-check.The manufacturer's product support engineer (pse) evaluated the device and confirmed the reported problem.During troubleshoot psa identified the issue was caused due to faulty flow sensor.To resolve the issue flow sensor assy.Was replaced by asp.System works as specified post this action.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.H11: updated contact information, contact office entity, and manufacturing site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key15890146
MDR Text Key305791633
Report Number2031642-2022-03008
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838025776
UDI-Public00884838025776
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberV60 VENT, CHINA OPT: CFLEX, AVAPS
Device Catalogue Number1076716
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/20/2023
Date Device Manufactured01/03/2017
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-