Model Number EL5ML
Device Problems Mechanical Problem (1384); Failure to Fire (2610); Noise, Audible (3273); Firing Problem (4011)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/22/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # x95c6l.The following additional information was obtained: "during a laparoscopic cholecystectomy, the clip feeding sound was unusual from the beginning.2 clips came out at the same time.When the device was fired inside the patient, a clip deployed properly, but there was an unusual feeling.The device was fired outside patient and 2 clips deployed at each firing." investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the el5ml device was returned without apparent damage.In addition, 4 partially formed clips and 2 conforming clips were returned inside a plastic bag.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled, and it fed and formed 3 conforming clips as expected.The jaws were examined for visual inspection and no anomalies were noted.No noise was heard.In order to evaluate the condition of the internal components, the device was disassembled, and no anomalies could be observed.Although no conclusion could be reached on the cause the malformed clips of the reported event, the instructions for use do contain the following caution: prior to loading a clip in the jaws and firing the instrument: ensure that the jaws are fully open by verifying that the line of demarcation between the jaws and the instrument shaft is past the distal end of the trocar cannula.Prior to positioning the jaws around the tubular structure or vessel, load a clip into the jaws by partially squeezing the trigger in a smooth continuous motion for approximately one-third of the total firing stroke.Position the jaws with the preloaded clip completely around the tubular structure or vessel to be ligated.The structure to be ligated should be positioned against the apex of the clip.Complete the firing cycle by squeezing the trigger until it stops against the handle to completely form the clip on the targeted structure or vessel.After firing, fully release the trigger.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot and batch number, and no non-conformance were identified.
 
Event Description
It was reported that during an unknown procedure, jamming occurred.Another device was used to complete the case.There were no adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key15894327
MDR Text Key307770971
Report Number3005075853-2022-08251
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberX95C6L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2022
Date Manufacturer Received11/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/16/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-