SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) Back to Search Results
Catalog Number C1032J
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/01/2022
Event Type  malfunction  
Event Description
It was reported that during the pre-use check, leakage of air from the product was observed.No patient injury reported.
 
Manufacturer Narrative
Udi and device serial number/lot number are unknown.No information has been provided to date.A product sample was received and is awaiting evaluation and investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
One (1) photo was attached to the complaint created.During the analysis conducted it could be observed a top view of circuit breathing product; however, no issues related to failure mode reported could be observed.The returned sample was visually inspected under normal conditions of illumination.No damages nor other workmanship defects were detected.The sample was submitted to a leak test inspection.The circuit was connected and successfully passed the test; thus, the failure mode reported was not confirmed.No root cause was determined, and no corrective actions were required since the complaint was not confirmed.No problem found.No lot number was provided; therefore, device history record review could not be completed.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
7-1-1 akasaka
minneapolis, MN 55442
MDR Report Key15895212
MDR Text Key307581842
Report Number3012307300-2022-27682
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC1032J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/16/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-