BAXTER HEALTHCARE CORPORATION MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number 5C4482 |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/09/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Initial reporter facility name: (b)(6).The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the twist clamp of a minicap transfer set could not be closed.This was observed during use of the device for peritoneal dialysis therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the device was received for evaluation.A visual inspection was performed with naked eye with no issues noted.Functional testing including leak, clear passage, and clamp function testing were performed with no issues noted.Torque engagement testing was performed, the engage and disengage force was acceptable and met specification.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
|
|