Model Number 302-20 |
Device Problems
Low impedance (2285); Electrical Shorting (2926)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/07/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that patient was seen in clinic for a programming adjustment.In clinic, the device showed low impedance.The patient did not feel any pain.X-rays have been completed.The patient recently had a generator replacement surgery a few months ago and it was noted that on the implant card the impedance was low, but still within normal limits at previous surgery.X-rays have not been reviewed by manufacturer to date.No known relevant surgical intervention has occurred to date.No other relevant information has been received to date.
|
|
Manufacturer Narrative
|
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova's employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects¿ or "malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
|
|
Event Description
|
Physician later reported that there was no trauma or manipulation prior to low impedance.X-rays were reviewed by manufacturer.The x-rays were provided with report of low impedance.The generator placement was located below rib 4.Based on the images provided, the feedthrough wires were intact, and the pin appeared to be fully inserted.There is strain relief present and it is placed per labeling.Three tie downs are seen and therefore are not placed per labeling, as only 2 should be used.There was a portion of the lead that appeared to be routed behind the generator.There was a small section of the lead in the neck that appeared as anomaly with the possibility of a fracture, however due to the quality of the image being lost when zooming in, the fracture/discontinuity cannot be confirmed.The cause of the patient¿s low impedance could not be determined based on the images provided.Note that the presence of a micro-fracture and/or a lead discontinuity could not be ruled out in the portion of the lead that is not visible in the provided images.No other relevant information has been received to date.
|
|
Search Alerts/Recalls
|
|
|