• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FT4 III; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FT4 III; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Model Number FT4 G3
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2022
Event Type  malfunction  
Manufacturer Narrative
The investigation excluded an interfering factor.For the differences in the values generated with the different analyzers, assays from different vendors can generate different values.This relates to the overall setups of the assays, the antibodies used, differences in reference materials/methods, and the standardization methodology used.From the information provided, a general reagent issue can most likely be excluded.The investigation did not identify a reagent problem.
 
Event Description
The initial reporter received questionable elecsys tsh v2 and elecsys ft4 iii assay results for 2 patient samples tested on the customer's cobas e 801 module with serial number: (b)(4), the investigation site's cobas e 411 module with serial number: (b)(4) and the investigation site's cobas e 801 module with serial number: (b)(4).This medwatch is for the ft4 iii assay.Refer to medwatch with patient identifier: (b)(6) for the tsh v2 assay.The exact date of the event is unknown.The date the information was received was used as the date of the event.The initial results were not reported outside of the laboratory.The ft4 iii reagent lot number used at the customer site's e 801 was requested but not provided.The ft4 iii reagent lot number used at the investigation site's e 411 is 593219 with an expiration date of 31-dec-2022.The ft4 iii reagent lot number used at the investigation site's e 801 is 630888 with an expiration date of 31-may-2023.Refer to the attachment in the medwatch for the highlighted questionable results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15916658
MDR Text Key307901987
Report Number1823260-2022-03863
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFT4 G3
Device Catalogue Number07976836190
Device Lot NumberASKU, 593219, 630888
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-