The hillrom technician found made three attempts requesting a resolution to this contact line without hearing back from the customer.Per the hillrom service manual the advanta¿ 2 should be subject to an effective maintenance program.Test each of the buttons to check that they activate the correct function and that they do not work intermittently by pressing each button for several seconds.Each movement must be continuous.Replace the pendant if necessary.As no serial number was provided for this bed, a search of the hillrom maintenance records for any hillrom performed preventative maintenance on this bed was unable to be completed.It is unknown if the facility performs preventative maintenance on their beds.Hillrom contacted the account three times trying to obtain the resolution for this event.No response has been received from the account.Based on this information, no further action is required.
|