Model Number UNKNOWN |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 11/06/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that on (b)(6) 2022, during the procedure, the surgeon placed two catheters bilaterally using tunnelers; the surgeon experienced difficulty placing one side.The catheters were removed on (b)(6) 2022 and the catheter which had been difficult to place broke during removal; there was no reported injury.Additional information received on (b)(6) 2022, the doctor reported, the patient was currently under palliative care and there was no plan to remove the tip of the catheter.The doctor additionally reported that he examined the catheter himself and it appeared that less than 5 mm was still possibly left inside if any was still left inside.The patient was due to be seen in the clinic in a week.
|
|
Manufacturer Narrative
|
The actual complaint product was not returned for evaluation.A review of the device history record is not possible as no lot number was provided.Root cause could not be determined.All information reasonably known as of (b)(6) 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to (b)(4), in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.Device not returned.
|
|
Manufacturer Narrative
|
The product involved in the report has been returned and is being processed for evaluation.All information reasonably known as of (b)(6) 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
|
Manufacturer Narrative
|
The actual sample from the reported event was returned for evaluation.Visual examination of the device revealed: the black tip was not present and the entire mid-body segment and infusion segment were stretched.The reported event could be confirmed as reported; broken tip; the root cause is undetermined.All information reasonably known as of 09 feb 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
|