Model Number PT-113950
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problems Failure of Implant (1924); Unspecified Infection (1930)
Event Date 07/19/2022
Event Type  Injury  
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2022-01922-1.0001825034-2022-01923-1.Concomitant medical products: item#: 113042, versa-dial 46x18x53 hum head; lot#: 190190, item#: 118001, versa-dial/comp ti std taper; lot#: 814790, item#: 113630, comp primary stem 10mm mini; lot#: 992200, item#: 113954, md hybrid glenoid base 4mm; lot#: 899530.During the investigation process a review of the sterile certifications were reviewed and found to be conforming with no applicable deviations.Devices were verified to have gone through acceptable sterilization process following iso/aami/astm & eu published guidelines.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet, such as external factors, i.E.Hospital/surgical environment, provider related risk factors, and/or patient comorbidities/risk factors.As there are no indications of a product or process issues identified affecting implant safety or effectiveness, therefore implanted products are not identified as the source or contributing to the reported infection.Visual examination of the provided picture identified that product has been explanted and is till assembled to mating implant.Part and lot numbers cannot be confirmed.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Radiographs were provided; however, they were not submitted for review as the images are in the immediate postop timeframe as reflected by the staple line present.Sending the images would not enhance the investigation.Images taken (b)(6) 2017.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that a patient underwent an initial right shoulder arthroplasty.Subsequently, approximately seven (7) years later, the patient was revised due to loosening of implants and an infection.All components were removed and replaced with an antibiotic cement spacer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PT HYBRID GLEN POST REGENEREX
Type of Device
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED/EXTREMITIES
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15946090
MDR Text Key305137932
Report Number0001825034-2022-02758
Device Sequence Number1
Product Code MBF
UDI-Device Identifier00880304459854
UDI-Public(01)00880304459854(17)270104(10)370430
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/08/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPT-113950
Device Catalogue NumberPT-113950
Device Lot Number370430
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/04/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age64 YR
Patient SexFemale
-
-