Model Number GWF-35-22015 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/18/2022 |
Event Type
malfunction
|
Event Description
|
The catheter kinked at 1,030 mm from the tip, so it became unusable and was replaced with a new catheter.The procedure was completed without any problems with the new guidewire.Additional/amended information - 29-nov-2022: core wire was stick out of the coil, and the guidewire was kinked at 1,030 mm from the tip.So, it became unusable and was replaced with a new guidewire.The procedure was completed without any problems with the new guidewire.
|
|
Manufacturer Narrative
|
No device was returned for analysis.As reported: core wire was stick out of the coil, and the guidewire was kinked at 1,030 mm from the tip.So, it became unusable and was replaced with a new guidewire.The procedure was completed without any problems with the new guidewire.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.At this time it is not possible to assign a definitive root cause for the event as reported.Based on the information provided by the supporting documentation "excessive force used", "material fatigue" and/or "improper use", may have impacted on the event as reported.
|
|
Event Description
|
The catheter kinked at 1,030 mm from the tip, so it became unusable and was replaced with a new catheter.The procedure was completed without any problems with the new guidewire.Additional/amended information - 29-nov-2022 core wire was stick out of the coil, and the guidewire was kinked at 1,030 mm from the tip.So, it became unusable and was replaced with a new guidewire.The procedure was completed without any problems with the new guidewire.
|
|
Manufacturer Narrative
|
The observation made during the investigation determined that the returned guidewire was bent on the main body.The j shape was under-formed.There was no evidence of the core wire sticking out from the coil when visually and microscopically examined.Both the distal and proximal joints were intact passing finger pull test, which confirmed guidewire was not fractured.A blood like substance was noted in the pouch and on the inner side of the dispenser tube, which would indicate that the guidewire was used in a medical procedure.No other damage was noted on the returned product.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.At this time, it is not possible to assign a definitive root cause for the event as reported.Based on the information provided by the supporting documentation "excessive force used", "improper use", "improper handling", "user failure in preparations for use" and/or "improper removal from packaging" may have impacted on the event as reported.
|
|
Event Description
|
The catheter kinked at 1,030 mm from the tip, so it became unusable and was replaced with a new catheter.The procedure was completed without any problems with the new guidewire.Additional/amended information - (b)(6) 2022 core wire was stick out of the coil, and the guidewire was kinked at 1,030 mm from the tip.So, it became unusable and was replaced with a new guidewire.The procedure was completed without any problems with the new guidewire.
|
|
Manufacturer Narrative
|
The observation made during the investigation determined that the returned guidewire was bent on the main body.The j shape was under-formed.There was no evidence of the core wire sticking out from the coil when visually and microscopically examined.Both the distal and proximal joints were intact passing finger pull test, which confirmed guidewire was not fractured.A blood like substance was noted in the pouch and on the inner side of the dispenser tube, which would indicate that the guidewire was used in a medical procedure.No other damage was noted on the returned product.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.At this time, it is not possible to assign a definitive root cause for the event as reported.Based on the information provided by the supporting documentation "excessive force used", "improper use", "improper handling", "user failure in preparations for use" and/or "improper removal from packaging" may have impacted on the event as reported.
|
|
Search Alerts/Recalls
|
|