The initial reporter stated they received questionable results for one patient sample tested with ck creatine kinase on two cobas 6000 c (501) module analyzers.No incorrect results were reported outside of the laboratory.The customer suspected the sample contains an interfering factor against a component of the ck assay.The sample was tested on the first c 501 analyzer, resulting in a value of < 6 u/l with a data flag.The sample was tested twice on the second c 501 analyzer, each time resulting in a value of < 6 u/l with a data flag.A serial number of (b)(4) was provided for one of the two c 501 analyzers.It is not known if this was the first analyzer or the second analyzer used for testing.The serial number of the other c 501 analyzer was requested, but not provided.
|