• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY (BD) PAXGENE® BLOOD RNA TUBE; RNA PREANALYTICAL SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY (BD) PAXGENE® BLOOD RNA TUBE; RNA PREANALYTICAL SYSTEMS Back to Search Results
Model Number 762165
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/16/2022
Event Type  malfunction  
Manufacturer Narrative
Bd had not received samples, but 2 photos were provided for investigation.The photos were reviewed and although the indicated failure mode for cracked tubes when frozen with the incident lot was observed, the bd recommendations for freezing were not followed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode cracked tubes when frozen.Bd was not able to identify a root cause for the indicated failure mode.
 
Event Description
It was reported when using the paxgene® blood rna tube there was a cracked tube.The following information was provided by the initial reporter.The customer stated: "the lab purchased those tubes on (b)(6) 2022 , they froze them and the tubes were sever cracked.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PAXGENE® BLOOD RNA TUBE
Type of Device
RNA PREANALYTICAL SYSTEMS
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK 
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
UK  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15960006
MDR Text Key307890554
Report Number9617032-2022-01257
Device Sequence Number1
Product Code NTW
UDI-Device Identifier54053727621653
UDI-Public54053727621653
Combination Product (y/n)N
Reporter Country CodeIS
PMA/PMN Number
K082150
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date03/31/2023
Device Model Number762165
Device Catalogue Number762165
Device Lot Number1167777
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-