MAUDE Adverse Event Report: PRISMATIK DENTALCRAFT, INC. COMFORT HARD-SOFT SPLINTTHERMOFORM MOUTHGUARD; MOUTHGUARD, PRESCRIPTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRISMATIK DENTALCRAFT, INC. COMFORT HARD-SOFT SPLINTTHERMOFORM MOUTHGUARD; MOUTHGUARD, PRESCRIPTION Back to Search Results
Device Problem Insufficient Information (3190)
Patient Problem Hypersensitivity/Allergic reaction (1907)
Event Type  Injury  
Manufacturer Narrative
The device has not been returned.If/when there is more information provided, a supplemental report will be submitted.Is not applicable with the exception of serial number as the device is manufactured by prescription.Is not applicable as the device is manufactured by prescription and not implantable.
 
Event Description
It was reported that the patient had a reaction to the comfort hard soft splint that was issued.It is unclear when the patient received the device, when the patient first used the device, or when the reaction occurred or resolved.However, it is noted that the patient experienced an abscess/ bump on gums.The device was worn for a few nights.There are no allergies noted.
 
Manufacturer Narrative
Capa 2023-006 the device has not been returned.However, the non-visual device evaluation has been completed and the results are as follows: dhr results no lot number provided, dhr not completed.Stock product reviewed results no stock product was available for review since the device was fabricated per physician's prescription only.Investigation methods/results the customer has not returned the complaint part for investigation to date.However, the non-visual device investigation has been completed.Root cause a root cause for this complaint cannot be explicitly determined.Ifu 9091 rev 4.0 (comfort h/s bite splint instruction for use) states "brush and floss your teeth before use.Rinse mouth well with clean water before inserting the device.If patient uses mouthwash, all traces of mouthwash should be removed by thoroughly rinsing out mouth with water.Rinse bite splint well with clean, cool water before and after use.Clean bite splint with clean, cool water only and let air dry." ifu provides warning "do not clean or soak in mouthwash; do not use denture cleanser, hot water, alcohol, hydrogen peroxide; do not place in direct sunlight".Glidewell research team and namsa conducted a series of testing on a similar thermoformed sleep device (haley) following iso 10993 (biological evaluation of medical devices) and the device was evaluated for potential cytotoxicity, skin irritation, delayed dermal contact sensitization and oral mucosal irritation.The haley test article was thermoformed with layers of erkodent material (erkoloc-pro and erkodur).The test results were listed below and summarized in biocompatibility report for haley sleep device (rpt 9733 rev 1.0) · for cytotoxicity testing, the test article extract showed no evidence of causing cell lysis or toxicity.· for skin irritation, there was no erythema and no edema observed on the skin of the animals treated with the test article.· for sensitization testing, the test article extracts showed no evidence of causing delayed dermal contact sensitization.· the test article showed nonirritant to the oral mucosa as compared to the control article.The device materials have been found to be biocompatible through the testing.There was no cytotoxic, sensitization, skin irritation, or oral mucosal irritation found in any of the test articles.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMFORT HARD-SOFT SPLINTTHERMOFORM MOUTHGUARD
Type of Device
MOUTHGUARD, PRESCRIPTION
Manufacturer (Section D)
PRISMATIK DENTALCRAFT, INC.
2144 michelson drive
irvine CA 92612
Manufacturer (Section G)
PRISMATIK DENTALCRAFT, INC.
2144 michelson drive
irvine CA 92612
Manufacturer Contact
herbert crane
2144 michelson drive
irvine, CA 92612
9495021907
MDR Report Key15963670
MDR Text Key305341213
Report Number3011649314-2022-00725
Device Sequence Number1
Product Code MQC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121365
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/12/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/03/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-