The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the heatsink assembly had been incorrectly installed.Additionally, the lamp provided a life meter reading of 100+ hours.The top cover was bent, and the front panel was discolored, and the chassis was found rusted.If additional information becomes available following the device evaluation, a supplemental report will be filed.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Please see the updates in sections h6 and h10.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely the heatsink assembly was rusted due to inappropriate heat sink installation.Olympus will continue to monitor field performance for this device.
|