H4: device manufacture date: march 7, 2022 - march 8, 2022.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition; there was no evidence of kink or bend observed on the tubing line from the photo.The tubing line appeared to be in normal condition, however, it was noted that the blue slide clamp was closed on the tubing line which would result in a no flow of fluid.The reported condition was verified.The cause of the condition could not be determined; however, the reported condition may potentially be due to a use error as during product use, the blue slide clamp must be opened to allow flow of fluid through the tubing line.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|