Model Number CI-1601-04 |
Device Problems
Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
|
Patient Problems
Erosion (1750); Unspecified Infection (1930); Tissue Breakdown (2681); Skin Infection (4544)
|
Event Date 05/16/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Event Description
|
The recipient is reportedly experiencing skin breakdown near the implant site.The recipient was hospitalized with an infection.A wound cleaning will be performed.The recipient is using the device.
|
|
Manufacturer Narrative
|
The recipient is currently not using the device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The recipient's infection was treated with antibiotics.Additional information regarding treatment details were not provided.The recipient's infection resolved and resumed device use.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.Due diligence attempts to obtain additional information regarding recipient's status were unsuccessful.The external visual inspection revealed the electrode was severed.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical tests performed.The device passed the mechanical test performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
The recipient reportedly experienced continued growth of dehiscence at the implant site.The recipient's device was explanted.The recipient was not re-implanted.The recipient is healing.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|