Model Number 40-8226 |
Device Problem
Break (1069)
|
Patient Problems
Pneumothorax (2012); Abdominal Distention (2601)
|
Event Date 11/19/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
The sample is reported to be available, but has not yet been received by the manufacturer.A review of the device history record is in-progress.All information reasonably known as of 09-dec-2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
|
Event Description
|
It was reported via fda medwatch/fda user facility report # mw (b)(4) the following information: nasojejunal (nj) tube clogged and was flushed with viokace to help resolve.Later that evening patient noted to have abdominal distention.X-ray revealed fractured tube and pneumoperitoneum.Patient went to or for repair of the bowel and removal [of] the nj tube." additional information received 22-nov-2022 stated the device was in place on day-1.Unfortunately, the patient required surgical intervention.".
|
|
Manufacturer Narrative
|
The device history record for the reported lot number, 30167129, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.All information reasonably known as of 12-jan-2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
|
Search Alerts/Recalls
|
|