• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO UPPER G.I.SCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO UPPER G.I.SCOPE Back to Search Results
Model Number EG29-I10
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2022
Event Type  malfunction  
Event Description
The patient went for gastroscopy due to stomachache, during use, a black shadow was found in the corner of the observation sight.This event occurred at the time of during use.There was no report of patient harm.
 
Manufacturer Narrative
If additional information becomes available, a supplemental report will be filed with the new information.
 
Manufacturer Narrative
Correction information: g6: follow up #1 h2:if follow-up, what type? h3:device evaluated by manufacture h6: coding changed based on the investigation result.D4:unique identifier (udi) d8:was this device serviced by a third party? h4: device manufacture date.Evaluation summary: based on the reported events, it is speculated that the endoscope was used without being dried after being reprocessed, causing water to enter from the tip of the endoscope and be reflected in the field of view.The scope was repaired by the third party and returned to the hospital.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO UPPER G.I.SCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key15974596
MDR Text Key308359875
Report Number9610877-2022-00671
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K131902
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEG29-I10
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/30/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-