• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH TELESCOPE, 30°, 4 MM; RIGID ENDOSCOPE TELESCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH TELESCOPE, 30°, 4 MM; RIGID ENDOSCOPE TELESCOPE Back to Search Results
Model Number A22002A
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/11/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned for evaluation and the customer¿s allegation was confirmed.It was noted that the outer tube was bent and the light guide post was damaged.The distal fiber was broken and the lens in the optical system was chipped.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The customer reported to olympus that the telescope, 30 degree, 4 mm had a bent light post that would not screw on.There were no reports of patient harm associated with this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely the damaged light post occurred due to the use of excessive force by the customer.However, a definitive root cause could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 30°, 4 MM
Type of Device
RIGID ENDOSCOPE TELESCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15982847
MDR Text Key308485491
Report Number9610773-2022-00653
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761020855
UDI-Public04042761020855
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K897003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22002A
Device Catalogue NumberA22002A
Device Lot Number412965
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2022
Was the Report Sent to FDA? No
Date Manufacturer Received05/04/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-