Model Number A22002A |
Device Problem
Material Twisted/Bent (2981)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/11/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned for evaluation and the customer¿s allegation was confirmed.It was noted that the outer tube was bent and the light guide post was damaged.The distal fiber was broken and the lens in the optical system was chipped.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
|
Event Description
|
The customer reported to olympus that the telescope, 30 degree, 4 mm had a bent light post that would not screw on.There were no reports of patient harm associated with this event.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely the damaged light post occurred due to the use of excessive force by the customer.However, a definitive root cause could not be determined.Olympus will continue to monitor field performance for this device.
|
|