Catalog Number V041163000000 |
Device Problems
Mechanical Problem (1384); Intermittent Energy Output (4025)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
There has been a previous report received where this malfunction has caused file separation.Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
|
Event Description
|
In this event it is reported that vdw.Silver reciproc stops during treatment.No injury occurred.
|
|
Manufacturer Narrative
|
Investigation: (b)(4):battery (voltage collapses under load) defective, micromotor smr1768215 + angle piece 24145(2011) (cannot be separated from each other, creates bent) defective, display pane (holding lug broken) defective, the housing is broken in several places (probably due to impact case) has no effect on the function.Foot control (b)(4)checked, no errors.Numbers of hours of use: 56:24:25.All complaints are monitored through the monthly product surveillance committee and a corrective action could be determined by the committee.
|
|