• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VDW GMBH VDW.SILVER RECIPROC; CONTROLLER, FOOT, HANDPIECE AND CORD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VDW GMBH VDW.SILVER RECIPROC; CONTROLLER, FOOT, HANDPIECE AND CORD Back to Search Results
Catalog Number V041163000000
Device Problems Mechanical Problem (1384); Intermittent Energy Output (4025)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where this malfunction has caused file separation.Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it is reported that vdw.Silver reciproc stops during treatment.No injury occurred.
 
Manufacturer Narrative
Investigation: (b)(4):battery (voltage collapses under load) defective, micromotor smr1768215 + angle piece 24145(2011) (cannot be separated from each other, creates bent) defective, display pane (holding lug broken) defective, the housing is broken in several places (probably due to impact case) has no effect on the function.Foot control (b)(4)checked, no errors.Numbers of hours of use: 56:24:25.All complaints are monitored through the monthly product surveillance committee and a corrective action could be determined by the committee.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VDW.SILVER RECIPROC
Type of Device
CONTROLLER, FOOT, HANDPIECE AND CORD
Manufacturer (Section D)
VDW GMBH
bayerwaldstrasse 15
munich bayern 81737
GM  81737
Manufacturer (Section G)
VDW GMBH
bayerwaldstrasse 15
munich bayern 81737
GM   81737
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key15987635
MDR Text Key306191556
Report Number9611053-2022-00692
Device Sequence Number1
Product Code EBW
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 01/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberV041163000000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date12/14/2022
Date Manufacturer Received12/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage A
Patient Sequence Number1
-
-