It was reported that the procedure was to treat a lesion in the left anterior descending (lad) artery with moderate calcification and moderate tortuosity.The 4.0x18mm xience sierra stent delivery system (sds) was noted to have a fracture between the metal and plastic casing [shaft] during the procedure but remained intact and resulted in the inability to deploy contrast to the balloon.The stent was partially deployed but balloon remained stuck to the internal structure of the stent.The balloon and sds were removed, however, the stent needed to be further expanded with the use of multiple balloons.There was no adverse patient sequela and there was no clinically significant delay in the procedure.No additional information was provided.
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents/complaints from this lot.The investigation was unable to determine a conclusive cause for the reported shaft break; however, reported difficult stent activation, difficult to remove and subsequent treatment appear to be related to the operational context of the procedure.In this case, it is possible that inadvertent mishandling and/or manipulation of the stent delivery system during the procedure may have contributed to the reported shaft break; however, this cannot be confirmed.Additionally, it was reported that the shaft break resulted in the inability to deploy contrast to the balloon, resulting in the reported difficult stent activation.The difficult stent activation likely contributed to an interaction between the balloon and deployed stent, resulting in the reported difficult to remove.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|