Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/16/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during a hip procedure, the threads of the shell inserter broke off inside the shell while trying to thread the shell.A new shell and inserter were opened to complete the surgery.There was no direct contact with the patient.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
(b)(4).Concomitant medical products: 010000668 ¿ g7 shell - 7152828.Product has been received and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified tip is broken.The shaft of the inserter has scuffing/scratching present.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional event information to report at this time.
|
|