The customer reports during a polypectomy procedure using a single use ligating device, the physician cut the loop of the ligation device to the end and severed the polyp.The excised polyp had insufficient resection area and the loop failed to resect the remaining polyp.The procedure was transferred to endoscopic mucosal resection (emr) and completed.There was no patient impact related to this event.According to the physician who used it, the stopper may have come off.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the reported event was likely caused by the following mechanism: 1.The polyp was surrounded with the loop.2.The polyp was ligated with the loop by pulling the slider.3.The slider was pulled excessively.This had caused the polyp to be severed mechanically.Furthermore, if the loop stopper had detached from the device, it was possible that the slider was pulled more after the polyp was severed.The event can be detected/prevented by following the instructions for use (ifu) which state: ¿do not apply unnecessary force to the loop when ligating tissue during the procedure.Excessive force applied to the loop could cut the surrounding body cavity tissue, resulting in patient injury, such as hemorrhages or mucous membrane damage.It may also damage the endoscope and/or instrument.¿ this supplemental report includes a correction to g2 to provide information that was inadvertently not included on the initial medwatch.Olympus will continue to monitor field performance for this device.
|