MAUDE Adverse Event Report: AGILENT TECHNOLOGIES SINGAPORE (INTL.) PTE LTD. AUTOSTAINER LINK 48; AUTOMATED SLIDE STAINER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGILENT TECHNOLOGIES SINGAPORE (INTL.) PTE LTD. AUTOSTAINER LINK 48; AUTOMATED SLIDE STAINER Back to Search Results
Model Number AS480
Device Problems Mechanical Problem (1384); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/22/2022
Event Type  malfunction  
Manufacturer Narrative
This malfunction may potentially impact staining performance.No erroneous staining result was reported by the customer in connection with this incident.No patient or user harm was indicated.Patient identifier-race: patient information has not been provided by the user.
 
Event Description
On (b)(6) 2022, the customer from germany reported dispensing issue.Some slides were not completely stained, the upper control was partly not stained.Sometimes there was too little reagent.The workflow was not affected.The field service engineer inspected the instrument and replaced the aspiration and dispense check valve which resolved this malfunction.The instrument is fully operational, within specification, and ready for the user.Diagnostics were not altered.No harm or patient impact was indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOSTAINER LINK 48
Type of Device
AUTOMATED SLIDE STAINER
Manufacturer (Section D)
AGILENT TECHNOLOGIES SINGAPORE (INTL.) PTE LTD.
no.1 yishun avenue 7
singapore north east, 76892 3
SN  768923
Manufacturer (Section G)
SHANDON DIAGNOSTICS LIMITED
tudor road
manor park
runcorn, WA7 1 TA
UK   WA7 1TA
Manufacturer Contact
mary o'neill
1834 state highway 71 west
cedar creek, TX 78612
3026338510
MDR Report Key16012780
MDR Text Key308201931
Report Number3003423869-2022-00135
Device Sequence Number1
Product Code KPA
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Repair
Type of Report Initial
Report Date 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAS480
Device Catalogue NumberAS48030
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/22/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/26/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-