It was reported that the doctor was placing an 6fr inlay ureteral stent after a ureteroscopy, but the stent would not advance over a solo flex 0.38 guidewire.After opening another 6fr inlay ureteral stent, the doctor was easily able to backload the stent into the patient over the same solo flex 0.38 guidewire.The determination was that the stent was defective.
|
The reported event is confirmed, cause unknown.The ureteral stent was returned without the original packaging.An evaluation was completed by futurematrix on 22sep2022.Unknown black flaking was present in the ziploc bag with the sample.The suture was observed to be tied and cut off, this is not the way the suture is placed on the stent during assembly.The tip of the proximal end of the stent appear to have notches in no certain pattern, as this was not reported by the customer and it is unknown if these notches were put in the stent by the user during insertion a new complaint will not be created.When inserting a 0.038" guidewire it could be confirmed that there was blockage.When continuing to push the guidewire through, a solid unknown plug appeared to be the cause of he blockage.The plug continued to flake and cause the same black flaking that was noted in the ziploc bag.The plug could not be identified with anything used at the biomerics fmi facility or from anything supplied with the stent for use.A potential root cause for this event could be, "part geometry".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "the insertion of a ureteral stent should only be done by those individuals who have comprehensive training in the techniques and risks of the procedure." "directions for use: determine the proper stent length for the patient.This is generally calculated from the baseline pyelogram.Accurate measurements will optimize drainage efficiency and patient comfort.Submerge stent in sterile water to activate the coating.Insert the cystoscope then pass the guidewire* through the scope until the tip is in the renal pelvis.Move the pigtail straightener over the proximal end (kidney coil end) of the ureteral stent allowing easier insertion onto the guidewire.Remove pigtail straightener once the stent is secure on the guidewire.Pass the stent over the guidewire through the cystoscope by using the push catheter for proper placement.Watch the distal end (bladder coil end) of the stent or the radiopaque, proximal end of the pusher.Stop advancing when the stent¿s distal end marker reaches the ureterovesical junction (uvj).**(see below for proper placement directions on the multi-length ureteral stent.) withdraw the guidewire slowly.The stent will form a pigtail automatically.Carefully remove the push catheter." the actual/suspected device was inspected.
|