The suspect analysis was judged "negative".The sysmex xs-1000i/xs-800i software guide (sg) chapter 6 - appendix, section 6.1 - ip messages, details the method in which the analyzer conveys its findings.Results without an error message are categorized as "positive" or "negative" based upon preset criteria, some of which are laboratory-defined.The system bases judgments on comprehensive surveys of numerical data, particle-size distributions, and scattergrams.Flags and messages, communicated through ip messages, indicate the analyzer's findings and notify of possible sample abnormalities.Any flag not generated indicates the criteria defined were not met.Investigation determined the analyzer performed as designed.The data suggests an inadequately mixed sample processed in the manual mode contributed to the event.No systemic deficiency was identified.When samples are inadequately mixed, rbcs settle toward the bottom of the tube, wbcs and platelets closer to the top, which can lead to the generation of erroneous results.The operator is responsible for proper collection and handling of samples and reporting results with repeat collection and testing when indicated.The sysmex xs-1000i instructions for use (ifu) chapter 3 - design and function, section 3.6 - analysis mode, provides instructions to the user in the manual mode: in manual mode, after mixing a sample manually, place the sample tube in the sample position without removing the cap.Chapter 6 - operation, section 6.11.3 - sample analysis, instructs the user: gently invert sample 10 times.Then place it into the sample position.
|