Model Number N/A |
Device Problem
High Test Results (2457)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/29/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The united states customer obtained discordant elevated atellica im folate results for two patients.The initial results were reported to the physician(s), who did not question the results. the same samples were repeated using folserum on the same atellica im analyzer, and the results were low.The interpretation of results section of the atellica im folate instructions for use states: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings." siemens continues to investigate.
|
|
Event Description
|
The customer obtained discordant elevated atellica im folate results for two patients.The initial results were reported to the physician(s), who did not question the results. the same samples were repeated using folserum on the same atellica im analyzer, and the results were low.There are no reports that treatment was altered or prescribed or adverse health consequences due to the discordant folate results.
|
|
Manufacturer Narrative
|
Initial mdr 1219913-2022-00452 was filed on 2022-dec-20 reporting an elevated result compared to repeat testing.Additional information on 2023-jan-09: siemens reviewed the information and determined that the elevated result is not attributable to the use of the incorrect fol calibration for serum samples because internal testing shows that use of the incorrect calibration yields lower values rather than higher values.Contributing factors such as pre-analytical factors and/or a sample handling/ stability issue cannot be ruled out.Internal siemens investigation indicates that the instrument and reagents were performing acceptably.The customer is operational.In section h6, the investigation finding, and investigation conclusion codes were updated.
|
|
Search Alerts/Recalls
|