ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME; DERMATOME HANDPIECE, ELECTRICALLY-POWERED
|
Back to Search Results |
|
Model Number N/A |
Device Problems
Failure to Shut Off (2939); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/08/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the device is working also when safe mode and sound is different.The event timing was during surgery and there is no harm or delay reported.Upon investigation, the device has unstable motor speed.No adverse events were reported as a result of this malfunction.Due diligence is complete.
|
|
Manufacturer Narrative
|
This complaint is recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the motor would operate without pressing switch and it produced unstable motor speed.The motor was replaced to resolve the reported issue.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Search Alerts/Recalls
|
|
|