While analysis was unable to be performed as the device was not returned, per the opinion of the physician, the root cause of the event was twiddling of the device by the patient.The device history record for this device serial number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Cvrx id# (b)(4).
|
A barostim system was implanted on (b)(6) 2022.During a routine follow-up on (b)(6) 2022, the device did not pass compliance and had low impedance.The patient reported experiencing strong pain in the ipg pocket several times since the start of therapy.X-rays were performed, and it was observed that the lead showed evidence of being twiddled, and the ipg had changed direction.It was noted that the ipg had not been sutured during the implant procedure.On (b)(6) 2022, a revision occurred.The pocket was opened, and the csl was untwisted and straightened.Thereafter, the impedance of the csl was in range and steady, and the system passed compliance.The ipg was sutured into the pocket.The patient was doing well following the procedure.
|