• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH RONG STRAIG W/4 L330; RONGEUR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH RONG STRAIG W/4 L330; RONGEUR Back to Search Results
Catalog Number 03.605.001
Device Problems Contamination /Decontamination Problem (2895); Compatibility Problem (2960)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2022
Event Type  malfunction  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Reporter is a j&j employee.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in germany as follows: during the cleaning process it was mentioned that the punches and rongeurs to a proper and successful cleaning, but it was not possible to eliminate visible contamination.The problem with the drum tongs and rongeurs is that they can be dismantled.Our drum tongs (rongeurs) and punches of this generation do not allow dismantling (dismantling for proper cleaning in the aemp).Despite repeated attempts with manual pre-cleaning, ultrasound cleaning and subsequent mechanical cleaning and disinfection, detectable blood residues remain on the surfaces via hemocheck-s.No patient harm.No surgery delay.This report is for one (1) rong straig w/4 l330.This is report 1 of 6 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6 investigation summary the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that rong straig w/4 l330 was observed with spots of foreign substance along the surface of the device.No signs of blood residues were identified.Since corresponding sterilization documents were not provided, it is not possible to establish a root cause of the observed condition.The important information (with cleaning and sterilization instructions) document was reviewed.Following relevant statements were found.Prior to steam sterilization, place the product in an approved sterilization wrap or container.The first and most important step in reprocessing all re-usable instruments is thorough (manual and/or mechanical) cleaning and rinsing.Thorough cleaning is a complex process whose success depends on various interrelated factors: water quality, quantity and type of cleaning agent, cleaning method (manual, ultrasonic bath, washer/disinfector), thorough rinsing and drying, proper product preparation, time, temperature, and thoroughness of the individual responsible for cleaning.All devices must be thoroughly cleaned and inspected prior to sterilization.Long, narrow lumens, blind holes, moving and intricate parts require particular attention during cleaning and inspection.During cleaning, only use cleaning agents that are labelled for use on medical devices and in accordance with the manufacturer¿s instructions (e.G.Temperature, contact time, and rinse time).Cleaning agents with a used dilution ph of within 7¿9.5 are recommended.Highly alkaline conditions (ph >11) can damage components/devices, such as aluminum materials.Do not use saline, environmental disinfection (including chlorine solutions) or surgical antiseptics (such as iodine- or chlorhexidine-containing products).Do not use a cleaning aid that can damage the surface of instruments such as steel wool, abrasive cleaners or wire brushes.Only place synthes devices with items of similar metallic composition together in an ultrasonic cleaner.Soiled or used synthes devices should not be loaded into a case for cleaning in a mechanical washer.Soiled synthes devices must be processed separate from trays and cases.A dimensional inspection for the rong straig w/4 l330 was not performed as it is not applicable to the complaint condition.A functional test was performed and the device was able to works correctly.The complaint condition was not able to be replicated.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the rong straig w/4 l330 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed? yes, reviewed.Dimensional inspection: n/a device history part number: 03.605.001 lot number: 528p365 manufacturing site: tuttlingen release to warehouse date: 01-dec-2021 a review of the device history records was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot part number: 03.605.001, lot number: 528p365, manufacturing site: tuttlingen, release to warehouse date: 01-dec-2021.A review of the device history records was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D10: date of concomitant therapy is (b)(6) 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RONG STRAIG W/4 L330
Type of Device
RONGEUR
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES TUTTLINGEN
unter hasslen 5
tuttlingen CO 78532
GM   78532
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16019057
MDR Text Key308446287
Report Number8030965-2022-11335
Device Sequence Number1
Product Code HTX
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.605.001
Device Lot Number528P365
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received02/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
LAMINECT-PUNCH 40°ANGL W/2 L330; LAMINECT-PUNCH 40°ANGL W/4 L330; LAMINECT-PUNCH 40°ANGL W/6 L330; RONG ANGL-UPWARDS W/3 L330; RONG STRAIG W/6 L330
-
-