Model Number INSULIN |
Device Problems
Incorrect Measurement (1383); Low Test Results (2458)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/01/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The sample was requested for investigation.
|
|
Event Description
|
The initial reporter stated they received questionable results for one patient sample tested with the elecsys insulin assay on a cobas 6000 e601 module, serial number (b)(4).The result did not agree with the patient's clinical manifestation and the values measured using a competitor method.The sample initially resulted in an insulin value of 11.31 uu/ml and repeated as 11.47 uu/ml on a second e601 module.The sample values were reported outside of the laboratory to the doctor and questioned.The sample was then repeated using a competitor method, resulting in an insulin value of 300 uu/ml.
|
|
Manufacturer Narrative
|
The last calibration with lot 590959 was performed on 22-sep-2022.Per product labeling: "renewed calibration is recommended as follows: after 1 month (28 days) when using the same reagent lot" quality control results were within specified ranges.The sample was provided for investigation.The customer's elecsys insulin result could be reproduced.Elecsys c-peptide was also tested and resulted in a similar value as obtained by the customer.Both values fit clinically together as within or close to the expected value range.The elecsys insulin and c-peptide results are considered correct.The investigation could not identify a product problem.Medwatch fields a2, a3 and b7 have been updated.
|
|
Search Alerts/Recalls
|