BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING Back to Search Results
Model Number 24657
Device Problems Difficult to Advance (2920); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2022
Event Type  malfunction  
Event Description
It was reported that the stent was difficult to advance, and when it was removed, it was damaged.An eluvia us, 6x60, 130 cm was selected for use in the superficial femoral artery (sfa) to treat peripheral artery disease via a right lower extremity angiogram.The device was prepped and inserted into the incision site.When the physician attempted to advance the eluvia delivery system through a previously placed stent, he reported feeling resistance.The eluvia delivery system was removed without issue.Once removed, it was observed that the tip of the stent was damaged.The eluvia was replaced with another eluvia in order to complete the procedure.The new stent was deployed without issue.There were no reported adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key16022754
MDR Text Key307588526
Report Number2124215-2022-52521
Device Sequence Number1
Product Code NIU
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P180011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number24657
Device Catalogue Number24657
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-