Model Number PE075F5
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/18/2022
Event Type  malfunction  
Event Description
It was reported that the balloon of a swan ganz catheter did not inflate before use.There were no patient complications reported.Per product evaluation findings, balloon did not inflate due to leakage from a tear and the balloon latex edges were not able to match up.
 
Manufacturer Narrative
The reported event of balloon inflation difficulties was confirmed.Balloon did not inflate due to leakage from a tear at proximal side of balloon latex.The balloon latex edges were not able to match up.No visible damage was found from catheter body, returned syringe and balloon windings.Further evaluation regarding related quality issues is under investigation.A device history record review was completed and documented that device met all specifications upon distribution.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Manufacturer Narrative
Engineering evaluation was completed and concluded a product risk assessment was met earlier.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN GANZ BIPOLAR PACING CATHETER FOR FEMORAL INSERTION
Type of Device
ELECTRODE, PACEMAKER, TEMPORARY
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
jonathan diaz
1 edwards way
irvine, CA 92614
MDR Report Key16024408
MDR Text Key307628008
Report Number2015691-2022-10209
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K813521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberPE075F5
Device Lot Number64419407
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/22/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-