Model Number N/A |
Device Problem
Power Problem (3010)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/20/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
This incident has been recorded under (b)(4).The results of the investigation are as follows: review of the most recent repair record identified the following related repair: tech confirmed the reported event was due to the motor speed being unstable.Tech replaced the motor, calibrated the unit, tested it, and returned it to the customer.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that outside of the surgery the device doesn't work due to low power.There was no injury, no additional medical intervention, no delay, and no conditions contributing to the event.No adverse event was reported as it relates to this event.No additional information available.
|
|