Model Number N/A
Device Problem Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/20/2022
Event Type  malfunction  
Manufacturer Narrative
This incident has been recorded under (b)(4).The results of the investigation are as follows: review of the most recent repair record identified the following related repair: tech confirmed the reported event was due to the motor speed being unstable.Tech replaced the motor, calibrated the unit, tested it, and returned it to the customer.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that outside of the surgery the device doesn't work due to low power.There was no injury, no additional medical intervention, no delay, and no conditions contributing to the event.No adverse event was reported as it relates to this event.No additional information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME, LINE-POWERED
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key16026854
MDR Text Key306410361
Report Number0001526350-2022-00839
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number31939200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2022
Was the Report Sent to FDA? No
Date Manufacturer Received05/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/09/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-