A steris service technician inspected the sterilizer and found that the ck4 valve was not operating properly.As the ck4 valve was not functioning properly this allowed water to back up in the sterilizer's chamber overnight.When the employee opened the door the next morning, steam emitted from the sterilizer subsequently causing the reported event to occur.The operator manual states (1-2), "steam may be released from the chamber when door is opened.Step back from the sterilizer each time the door is opened to minimize contact with steam vapor." the steris service technician counseled user facility personnel on the proper use and operation for the amsco 400 sterilizer.The technician rebuilt the ck4 valve, tested the sterilizer and confirmed the unit to be operating properly.The sterilizer was returned to service and no additional issues have been reported.
|