NEUROPACE, INC. NEUROPACE RNS SYSTEM Back to Search Results
Model Number RNS 320
Device Problems Failure to Interrogate (1332); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2022
Event Type  malfunction  
Event Description
On (b)(6) 2022, there was a report of an unresponsive neurostimulator after undergoing a mri.The device was placed in mri mode prior to the scan without complication.Once the scan was completed, the device would not establish telemetry.On (b)(6) 2022 the neurostimulator was replaced.The neurostimulator was returned and investigated by neuropace.The recently implanted replacement neurostimulator is programmed for detection and stimulation and is functioning as expected.
 
Manufacturer Narrative
(b)(4).Investigation results - confirmed the neurostimulator reset.This is consistent with a rare but understood series of events wherein an externally impressed voltage such as from mri or esu can interact with the device's internal regulated power supplies bringing them to zero volts.Though physically not damaged, the component does not recover from this state without a power cycle.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
6502382788
MDR Report Key16034786
MDR Text Key306173654
Report Number3004426659-2022-00051
Device Sequence Number1
Product Code PFN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberRNS 320
Device Catalogue Number1007927
Device Lot Number30601-1-1-1
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/30/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age34 YR
Patient SexFemale
-
-