• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD SURGIDAC; ENDOSCOPIC TISSUE APPROXIMATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD SURGIDAC; ENDOSCOPIC TISSUE APPROXIMATION DEVICE Back to Search Results
Model Number 170044
Device Problems Break (1069); Detachment of Device or Device Component (2907); Human-Device Interface Problem (2949)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2022
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: 173016 - 173016 endo stitch instrument, lot #: j2g0204ey.173016 - 173016 endo stitch instrument, lot #: j2g0204ey.173016 - 173016 endo stitch instrument, lot #: j2g0204ey.170044 -170044 endo stitch tri 2-0 grn 120 s'dac, lot #: j2b2352y.170044 -170044 endo stitch tri 2-0 grn 120 s'dac, lot #: j2b2352y 170044 - 170044.Endo stitch tri 2-0 grn 120 s'dac, lot#: unknown.170044 - 170044 endo stitch tri 2-0 grn 120 s'dac, lot #: unknown.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, on a laparoscopic roux-en-y gastric bypass and hiatal hernia repair, during the creation of side-to-side anastomosis, the devices had unloading difficulties.The needles broke, disengaged from the handle, and fell into the patients cavity.The needles were retrieved with a laparoscopic grasper.A single-use suturing device was used instead.There was no patient injury.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
According to the reporter, on a laparoscopic roux-en-y gastric bypass and hiatal hernia repair, during the creation of side-to-side anastomosis, the devices had unloading difficulties.Three needles broke, disengaged from the handle, and fell into the patient¿s cavity.All three needles were retrieved with a laparoscopic grasper.A single-use suturing device was used instead.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURGIDAC
Type of Device
ENDOSCOPIC TISSUE APPROXIMATION DEVICE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key16040561
MDR Text Key307732188
Report Number9612501-2022-02238
Device Sequence Number1
Product Code OCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K934738
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number170044
Device Catalogue Number170044
Device Lot NumberJ2B2352Y
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/05/2023
Date Device Manufactured02/22/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SEE H10.
Patient Age56 YR
Patient SexFemale
-
-