• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH DOBBHOFF 10FR 43IN W STYLET EN; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH DOBBHOFF 10FR 43IN W STYLET EN; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS Back to Search Results
Model Number 8884711006E
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/19/2022
Event Type  malfunction  
Event Description
The customer reported that the connector tip is detached from the feeding tube.The feeding tube is still in the patient.
 
Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Manufacturer Narrative
The device history record (dhr) was reviewed, and no abnormal process conditions were present during the manufacturing of the product that could have led to the reported issue.The dhr review showed that all acceptance criteria inspections per established sampling levels were within acceptable limits during the production process.An investigation was carried out with the multifunctional team.All processes and controls were found to be followed correctly, including packaging and all inspections performed on the product.No abnormal conditions were found that could trigger the reported issue.One (1) photo was provided for evaluation.Through the photo, the reported issue was observed; the yport was detached.A corrective and preventive action has been initiated to address the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOBBHOFF 10FR 43IN W STYLET EN
Type of Device
GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key16042899
MDR Text Key307774965
Report Number9612030-2022-03499
Device Sequence Number1
Product Code PIF
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884711006E
Device Catalogue Number8884711006E
Device Lot Number2022314964
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/21/2022
Date Device Manufactured08/19/2020
Is the Device Single Use? Yes
Type of Device Usage A
Patient Sequence Number1
-
-