Model Number P7900B000022 |
Device Problem
No Audible Alarm (1019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/28/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
No further information is available on the repair of the bed at this time.The investigation is ongoing, however if any additional relevant information is identified following completion of the investigation, the additional relevant information will be submitted in a supplemental report.
|
|
Event Description
|
Hillrom received a report from a hillrom technician stating the bed had no bed exit alarm.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint#: (b)(4).
|
|
Event Description
|
Hillrom received a report from a hillrom technician stating the bed had no bed exit alarm.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
|
|
Manufacturer Narrative
|
The hillrom technician found both fuds needed to be replaced.Per the hillrom service manual, it is necessary for the centrella¿ bed to have an effective maintenance program.We recommend that you do annual preventative maintenance (pm).Make sure the bed exit system operates correctly.Replace parts if necessary.A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed on january 2022.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced both fuds to resolve the reported event.Based on this information, no further action is required.
|
|