Catalog Number 8065977763
Device Problems Break (1069); Device Damaged by Another Device (2915); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2022
Event Type  malfunction  
Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed, and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported that during a cataract extraction with intraocular lens (iol) implant procedure, when the doctor implanted the lens in the patient's eye, it seemed like it was difficult to slide the lens through the cartridge.The cartridge broke at the tip and scratched the lens; the doctor didn't realize it until the lens unfolded inside the eye and was seen to be scratched from the center of the lens.The doctor decided to remove the lens.She broke it with tweezers and removed it.The circulating nurse provided the new iol to the doctor so that she could load it again and implant it; this lens was able to be implanted correctly by the doctor without any setbacks.Additional information has been requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16059283
MDR Text Key307583094
Report Number1119421-2022-02748
Device Sequence Number1
Product Code MSS
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
K063155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065977763
Device Lot Number15407815
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/02/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/27/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ACRYSOF IQ VIVITY TORIC EXTENDED VISION IOL; DISCOVISC
-
-