The user facility reported to olympus that during use of the hf- resection electrode, loop, 24 fr., 0,2 wire the device broke inside of the patient.The physician managed to removed it from the patient, and another device was opened to complete the procedure.Additionally, the procedure was delayed by 30 minutes.Attempts for additional information are in progress.
|
This supplemental report was submitted to provide the results of the legal manufacturer¿s investigation.Olympus followed up with the user facility to obtain additional information regarding the reported event but no information was obtained.The dhr review showed there is no non-conformity associated with this device with respect to the described issue and the device was manufactured according to valid instructions and met all specifications.The most probable cause of the reported issue is most likely due to wear and tear.It should be mentioned that the loop at the distal end of the electrode wears out during use and may break, burn or melt.In general, the customer is required to check the function of all devices used prior to a procedure.Additionally, according to the ifu, a suitable replacement device must be provided during an application.Olympus will continue to monitor complaints for this device.
|