MAUDE Adverse Event Report: HILL-ROM BATESVILLE CENTRELLA MED-SURG; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE CENTRELLA MED-SURG; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Model Number P7900B100020
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/02/2022
Event Type  malfunction  
Manufacturer Narrative
The hillrom technician found the siderail cable to the junction box needed to be replaced.Per the hillrom service manual, it is necessary for the centrella¿ bed to have an effective maintenance program.We recommend that you do annual preventive maintenance.Do the function checks as applicable for your bed.If the bed passes all of the checks for its configuration, do the necessary administrative tasks, and prepare the bed to be put into service.If the bed does not pass all of the checks, repair or replace the part as applicable.Do not put the bed into service until it passes all of the checks.Warning¿report to bed authorized maintenance persons any unusual sounds, burning odors, or movement deviations observed in the controls, motors, or limit switch functions.Check the cable connection between the mcb and bcb.Replace or connect the cable as necessary.Replace the bcb.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the siderail cable to the junction box to resolve the reported event.¿¿based on this information, no further action is required.
 
Event Description
Hillrom received a report from a hillrom technician stating the bed had error message "communication lost" and the bed sometimes moves on its own without pressing any buttons.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRELLA MED-SURG
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
brad wheeler
1069 state route 46 east
batesville, IN 47006
3128199307
MDR Report Key16063453
MDR Text Key308242043
Report Number1824206-2022-00546
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP7900B100020
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/02/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/30/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-