• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PIVOT GUARDIAN DIRECT FINE TRACTION; APPARATUS, TRACTION, NON-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PIVOT GUARDIAN DIRECT FINE TRACTION; APPARATUS, TRACTION, NON-POWERED Back to Search Results
Model Number 3105000306
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
Patient Problems Hip Fracture (2349); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/14/2022
Event Type  Injury  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that there was a patient injury of a blown out wall of the patient's femur.
 
Event Description
It was reported that there was a patient injury of a blown out wall of the patient's femur.
 
Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: patient injury of a blown out wall of femur.Probable root cause: design: poor connector designs or tolerating.Connector is loose.Connector unable to hold force.Unclear feedback if mechanism is fully locked/attached.Casters slip on floor.Vertical spars collapse.Horizontal spars collapse.Fine traction element pivots unintentionally.Boot disconnects from fine traction element.Tabletop extension cushion lifts or slides under force use error.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.H3 other text : 81.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIVOT GUARDIAN DIRECT FINE TRACTION
Type of Device
APPARATUS, TRACTION, NON-POWERED
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
andrea zenere
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key16064388
MDR Text Key306297933
Report Number0002936485-2022-00800
Device Sequence Number1
Product Code HST
UDI-Device Identifier07613327565379
UDI-Public07613327565379
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3105000306
Device Catalogue Number3105000306
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-