Model Number 3105000306 |
Device Problems
Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
|
Patient Problems
Hip Fracture (2349); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/14/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
|
Event Description
|
It was reported that there was a patient injury of a blown out wall of the patient's femur.
|
|
Event Description
|
It was reported that there was a patient injury of a blown out wall of the patient's femur.
|
|
Manufacturer Narrative
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: patient injury of a blown out wall of femur.Probable root cause: design: poor connector designs or tolerating.Connector is loose.Connector unable to hold force.Unclear feedback if mechanism is fully locked/attached.Casters slip on floor.Vertical spars collapse.Horizontal spars collapse.Fine traction element pivots unintentionally.Boot disconnects from fine traction element.Tabletop extension cushion lifts or slides under force use error.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.H3 other text : 81.
|
|