Three endobronchial valves (ebv) were placed in various left sided segments to treat a bronchopleural fistula (bpf) and alveolar-pleural fistula with persistent air leak (pal) following left upper lobectomy.More than 2 years post spiration valve placement, patient presented with dyspnea on exertion, occasional small-volume hemoptysis as well as significant weight loss in the previous 6 months.A ct scan showed a pneumothorax; one ebv remained well positioned in a segmental bronchus, whereas the other 2 appeared to have migrated into the pleural space.The two valves that were found to have passed into pleural space were removed.This was an off-label use.Under this indication for use (treatment of air leak) the valves are to be removed after six weeks.
|
Article discovered during literature search review: "j bronchol intervent pulmonol volume 24, number 2, april 2017; doi: 10.1097/lbr.0000000000000352; trevor m.Taylor, md, robert j.Lentz, md, fabien maldonado, md, fccp" no dates of implant or extraction were provided in the article, which was published in april 2017.Valve sizes also not provided.This was an off-label use.Under this indication for use (treatment of air leak) the valves are to be removed after six weeks.Pneumothorax is the most common device related serious adverse event that is associated with the spiration valve system.In the emprove study, the incidence of serious pneumothorax was 14.2%, with over 75% of the serious pneumothorax events occurring within the first 3 days post-procedure.Early-onset pneumothorax in the treatment group likely resulted from lung conformation changes due to acute reduction in lung volume by valve therapy, triggering rapid expansion of the ipsilateral non-targeted lobe leading to pneumothorax.(criner gj, delage a, voelker k, et al.Improving lung function in severe heterogenous emphysema with the spiration® valve system (emprove): a multicenter, open-label, randomized control trial.Am j respir crit care med.2019;200(11):1354-1362.Doi:10.1164/rccm.201902-0383oc).However, it should be noted that pneumothorax events are also recognized as an indicator of successful target lobe occlusion and when managed according to published expert guidelines (valipour a, slebos dj, de oliveira hg, et al.Expert statement: pneumothorax associated with endoscopic valve therapy for emphysema -- potential mechanisms, treatment algorithm and case examples.Respiration 2004 2014;87(6): 513-521.Doi:10.1159/000360642), subjects with pneumothorax events experienced clinical benefits similar to that in subjects without pneumothorax events (criner gj, delage a, voelker k, et al.Improving lung function in severe heterogenous emphysema with the spiration valve system (emprove).A multicenter, open-label, randomized control trial.Am j respir crit care med.2019;200(11):1354-1362.Doi:10.1164/rccm.201902-0383oc).The reported event aligns with the experience in the emprove clinical trial and is an expected adverse event associated with the spiration valve system.
|