The reported field event of missing silicone seal was verified.However, the issue was found to be consistent with having occurred during the procedure.Septum material was observed inside the lv (is-1) set screw hex cavity.This did not allow the set screw to be tightened and secured properly in the field.The manufacturing report showed sufficient adhesive traces and septum residues on the inner wall of the cavity.The device manufacturing history record was reviewed and showed that the product passed all quality control tests prior to its distribution.The device was above elective replacement indicator (eri) when received.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.
|