: h4: the device was manufactured from april 20, 2021 - april 27, 2021.H10: the actual device was received for evaluation.A visual inspection was performed using the naked eye found fluid inside the bag that contained the sample.Functional testing was performed by filling the device with green colored water.During fill, no evidence of leak was observed.After fill, the blue winged luer cap was used to close the distal luer end of the device (the non-baxter product was not returned with the sample).The sample was monitored until the next day and no signs of leak were observed.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|