Catalog Number 2C1009K |
Device Problem
Filling Problem (1233)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/06/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that a large volume infusor underinfused during patient use; the flow rate was found to be unusually slow.The expected infusion time was 48 hours; however, when the 48 hours was up, 100ml remained in the device.The device had been filled with fluorouracil diluted in 0.9% normal saline solution.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
Additional information was added to h3, h4, h6 and h10: h4: the device was manufactured from april 28, 2022 - april 30, 2022.The actual device was not available; however, a photograph of the sample was provided for evaluation.The photograph was visually inspected and showed fluid inside the bladder which suggested an underinfusion problem may have occurred.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|