• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem Device Contamination with Body Fluid (2317)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/22/2019
Event Type  malfunction  
Manufacturer Narrative
A getinge service territory manager (stm) was dispatched to investigate.The stm observed blood on the blood return/drain line up to the filter before the purge assembly.The safety disk and condensation removal module (crm) were also contaminated with blood and the fill line had tiny specks of blood.The stm replaced the blood detect tubing, safety disk assembly, condensation removal module (crm) assembly and clear polyurethane tubing.The stm noted that nothing unusual was identified in the iabp log files then completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.Patient or person height (cm) 165.This report is being submitted as the result of a retrospective review conducted in capa (b)(4).
 
Event Description
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) experienced a sudden severe intra-aortic balloon leak allowing blood to fill the helium tubing all the way to the safety disk.The iabp unit was turned off immediately.There was no known harm or injury to the patient and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key16075554
MDR Text Key308474386
Report Number2249723-2022-03429
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/22/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/20/2011
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
SENSATION PLUS 50CC BALLOON
Patient Age68 YR
Patient SexFemale
Patient Weight125 KG
-
-