DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Device Contamination with Body Fluid (2317)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/22/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge service territory manager (stm) was dispatched to investigate.The stm observed blood on the blood return/drain line up to the filter before the purge assembly.The safety disk and condensation removal module (crm) were also contaminated with blood and the fill line had tiny specks of blood.The stm replaced the blood detect tubing, safety disk assembly, condensation removal module (crm) assembly and clear polyurethane tubing.The stm noted that nothing unusual was identified in the iabp log files then completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.Patient or person height (cm) 165.This report is being submitted as the result of a retrospective review conducted in capa (b)(4).
|
|
Event Description
|
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) experienced a sudden severe intra-aortic balloon leak allowing blood to fill the helium tubing all the way to the safety disk.The iabp unit was turned off immediately.There was no known harm or injury to the patient and no adverse event was reported.
|
|